Technical Paper

Total Organic Carbon Analyzer

1991-07-01
911434
Development and testing of a high sensitivity monitor for the measurement of Total Organic Carbon (TOC) in water without gravity-dependent components and using minimal chemical reagents has been performed. A breadboard instrument was constructed and tested for linearity, selectivity, and day-to-day reproducibility. The instrument has a linear response for a wide range of organic compounds over a range from 20 ppb C to 50 ppm C, with excellent day-to-day reproducibility. The upper level can be extended to ∼100 ppm C by changing the operating conditions of the analyzer. The selectivity of the instrument has been determined and no interference is observed except for high concentrations (> 10 ppm) of iodine, hypochlorous acid, sodium nitrite and sodium sulfide.